Commissioning with Terminal PR 5610 Installation Manual
120 Sartorius
5.4.1.1 Calling up and display of the weighing point calibration mask
Calling up weighing point calibration mask
Calib/Config
Weighing point selection is by means of the cursor keys [] and []. The actually selected weighing
point is highlighted by inverse display.
Calling up the weighing point calibration mask is done by pressing key [C]
Calib/Config.
If a
production is running, this call is disabled, message:
WP is in use
When calling up mask
Calib/Config
with the CAL-switch open, tare memory and zero setting
memory are erased. Moreover, the following functions are disabled during weighing point
calibration: test, automatic zero tracking, taring and zero setting
The weighing point calibration mask is shown below:
PR1713 Rel. 03.12 20.05.2004 09:05
CALIBRATION Mo 05.July 2004 09:31:51
033.323
A < > 033.3kg
Parameters
Ana.Filter: on Meastime : 2.00 Standst.rn: 1.00 d Overload : 9 d
Dig.Filter: off Fcut : 1.00 Zeroset.rn: 50.00 d no p below: 50 d
Multirngmd: off Standst.ti: 0 Zerotrk.rn: 0.25 d Stepwidth : 1
Test mode :Absolute Zerotrk.ti: 0 Zerotrk.st: 0.25 d Multirang1: 3000 d
W and M : none Tare timot: 8 Supply : 12 Volt Multirang2: 6000 d
Current calibration
Fullscale: 600.0kg Deadload :-0.027408 mV/V Resolution: 8.774623 cd/c
Calipoint:---fixed--- Span : +0.889043 mV/V Sensitivity: 4.000000 µV/d
Keys
Select item with cursor up/down/left/right Change with +/-
Return to FiXed,Set Fullscale,Calibrate by Weight or Millivolts
E Exit calibration
The weighing point calibration mask contains all weighing point parameters. They are classified as follows:
Configuration parameter
Parameters
:
During configuration, the parameters which are not included into conversion of measured counts into a weight value
are determined for operation of the analog part.
Calibration parameters
Current calibration.
During calibration, the conversion of the analog input signal (or
digital: of measured counts) into the weight value is determined.
The access functions (
Keys
) are displayed below these two weighing point parameter groups.