Hardware Configuration PR 5610 Installation Manual
36 Sartorius
3.4.3 PR 1713/06 Analog output card
Connecting method: 2-row 25-pin connector
Terminal block A with 8 screw
terminals
Number of outputs:
1 current output, voltage by use
of external resistor
Output:
Gross, net weight or application
defined
Range: 0/4 ... 20mA, configurable via
software
Resolution: 16 bits binary, 20,000 internal
counts represent e.g. 0 - 20 mA
Linearity error: 0 - 20mA: 0.04 %
4 - 20 mA: 0.02 %
Temperature error: < 100 ppm/K
Error on zero: 0.05 %
Error on FSD: < 0.1 %
Load: 0 ... 500 Ohm max.
Protected against short circ.: yes
Dimension (LxWxH): 80 x 68 x 15 mm Potential isolation: yes
Weight: 33 grs. Cable length (screened): 150 m (current output)
After installation of the card a COLD start has to be performed.
The instrument can work with only 1 analog output (power consumption).
Set-up in [Config] of the application program is necessary. After cold start, the card is detected and the output
is set to 4...20 mA and gross weight.
Pin allocation of option PR 1713/06 (Installation see chapter 3.2.1)
Terminal block A analog A1 A2 A3 A4 A5 A6 A7 A8
lower row
output I +
out
I -
GND
(ext.)
nc nc nc nc nc nc