Commissioning PR 5610 Installation Manual
98 Sartorius
After [Delete] a 7 digit licence number can be entered and con-
firmed with . For a valid number the message 'PR17XX/XX
disabled' is displayed. Otherwise the message 'wrong Licence
number' is displayed. Leave the menu with
.
Prompt at leaving the menu.
Select [Save] for storing changes in the licence setup.
Select [noSave] if changes are temporary (only in RAM).
Select [Undo], if no changes have been made.
This safety prompt appearsm, if [Undo] has been entered.
With [YES] no changes are stored, with [NO] changes shall
become valid.
4.8.6 Show board number
In menu [Setup]-[Show Boardnumber] the 9-digit instrument specific board number (unique) is displayed.
Select the access with
.
The board number is required to order function or application
licences. Leave the menu with
.
4.8.7 Print setup data
Select access with . Print setup data via the selected
interface in [Setup]-[Serial Ports]-[Printer device at].
Error message if no interface was assigned to the printer device.
Leave the menu with
.
4.8.8 Last fault
Select access with . Print the last fault via the selected
interface in [Setup]-[Serial Ports]-[Printer device at].
Error message if no interface was assigned to the printer device.
Leave the menu with
.
4.8.9 Refresh display
Refreshing the weight display intensity.
The weight display elements, which are used very rarely, gradually lose their brightness. Consequently, the
luminosity e.g. within a digit can be different. However, regeneration of the display intensity is very easy.
Refreshing the display is started after pressing
. All display
segments are switched on for refreshing intensity, wait approx.
5 min. Termination of refresh with
.