Commissioning with Terminal PR 5610 Installation Manual
134 Sartorius
5.6.8 Error messages in the calibration menu
no standstill no standstill was reached during calibration.
value is invalid the parameter is too high or too low. Proposal of the highest
permissible value with: setting max. ...
ADC Error general error ADC/ load cells
current load is below deadload occurs when entering the SPAN value. The deadload was
removed. Subsequently, SPAN is below zero.
Current load is above full scale occurs when entering the SPAN value or the calibration
weight value. The input voltage or the input voltage
calculated for the FSD >FSD.
Illegal stepwidth the entered step width is not correct.
FSD not multiple of stepwidth FSD not divisible by stepwidth/ incompatible stepwidth.
Fullscale is below calibration point after calibration, a new FSD was entered again, which is
below the initial calibration point/ FSD.
Caution: subsequent entry of a larger scale is permissible,
without recalibration!
overflow in converting units These messages are only generated with different
dimensions.
underflow converting units To generate an error message, the scale must be incalculable,
e.g. FSD 1000 kg, entered calibration weight 1g (should be 1
t, which would be accepted).
Input voltage above max The actual or the calculated input voltage is > 36 mV
not enough counts per d The measurement signal for the specified scale is too small
device not ready
protocol error
BCC error
EAROM error
" "
" "
Canceled Cancel by pressing key Q