Commissioning PR 5610 Installation Manual
64 Sartorius
4.3.1.3 Parameter tree
Setup
- Set Clock
Time and Date
- Serial Ports
- Operator device at
<Builtin RS232>, none, Slot 1 .. 3 RS485, Slot 1 ... 3 RS 232 (select)
- Printer device at
<none>, Builtin RS232, Slot 1 .. 3 RS485, Slot 1 ... 3 RS 232 (select)
- Remote device at
<none>, Builtin RS232, Slot 1 .. 3 RS485, Slot 1 ... 3 RS 232 (select)
- Serial port setup
* Builtin RS232
Protocol, Baudrate, Bits, Parity, Stopbits, Devtype, Echo * if not sel. for operator dev.
* Slot 2 RS 485
Protocol, Baudrate, Bits, Parity, Stopbits, Devtype, Echo * if installed
- Software Parameter
- Language
<English>, Local Language
- Frontkey timeout
1 s, <2 s>, 3 s
- Low battery check
<on>, 1min, off
- Report to
<none>, communication, application, communic. & appl., application +Prod.
- Tare key
<enabled>, disabled
- Set zero key
<enabled>, disabled
- Quit in mainlevel
<enabled>, disabled
- Reset on stop+exit
<1 s>, 5 s, disabled
- S88.01 Interface
<off>, on
- Software download
<protected by setup>, always enabled
- Lines per recipe
1, 2, ...<10> ... 255
- Recipe simulation
<enabled>, disabled
- Subrecipe
enabled, <disabled>
- * Fieldbus Parameter
* only if PR 1721/XX is installed in slot 4
- PR 1721 protocol
<none>, Interbus-S, Profibus-DP, DeviceNet, CanBus-Mas
see chapter 4.8.4
- Show Boardnumber
Display: Boardnumber required for licences
- Licence setup
Show, Add, Delete
- Print Setupdata
Activate printing on the selected interface, failure: no printer device
- Refresh Display
All segments are illuminated
- I/O Slots
Display: Slot 1 ...4, card type or empty, @ task no.
- Show Version
Display: Firmware version, Application, BIOS version
- * Enable download
Download enabled * only if [Software download] is set to [protected by setup]
- Reboot
- Cold
Cold start
- Warm
Warm start
- Bios
If selected, no Warm start possible
- Cold
Cold start
- Erase
Default settings
- Flash
Load new software
- Test
Test ROM, RAM, EAROM, keypad, display