PR 5610 Installation Manual Commissioning
Sartorius 63
4.3.1.2 Internal ADC calibration tree
WP X: Internal ADC
- Assign
internal, liquid, user, GWT-Scale, x-BPIScale
- Calib
- New
- Reset all to default
YES or NO
- Set Fullscale
0.000100 ...<3000>... 9999900 <kg>, t, lb, g
- Set Stepwidth
<1>, 2, 5, select with [More] 10, 20, 50 , current stepwidth in ><
- Set Deadload
[Load] or [mV/V]
Enter Deadload
<0.000000 mV/V>
- Set Span
[Load] or [mV/V]
Enter current load
0.0001 ... 9999900 kg, t, lb, g
Enter Span
<1.000000 mV/V>
- Calculate test
Test: Calculate testfigure
- Exit calibration
YES [Save] or [Undo], NO
- Modify
Only for small modifications, in other cases use [New]
-
see
New
- Param
- Analog filter
<on>, off
- Digital filter
<off>, bessel, aperiod., butterw., tscheby.
- Multirange mode
<off>, on
- Test mode
<Absolute>, Relative
- W & M
<none>, OIML, NTEP, NSC
- * Fcut
0.10 Hz ... 5.00 Hz *only if Digital filter not off
- * Measuretime
0.05 ... <0.30 s> ... 2.00 s *only if Digital filter = off, else M = 0.05 s
- Standstill time
<1 M> ... 8 M
- Zerotrack time
<0 M> ... 100 M
- Tare timeout
0 M ... <8 M> ... 100 M
- Standstill range
0.00 d ... <1.00 d> ... 10.00 d
- Zeroset range
0.00 d ... <50.00 d> ... 500.00 d
- Zerotrack range
0.00 d ... <0.25 d> ... 500.00 d
- Zerotrack step
0.00 d ... <0.25 d> ... 10.00 d
- * Supply
<12 Volt>, 20 Volt * only for PR 5710
- Overload
0 d ... <9 d> ... 9999999 d
- Don't print below
0 d ... <50 d> ... 9999999 d
- * Multirange limit 1
0 d ... 9999999 d, crossover from small to mean stepwidth
*only if Multirange mode = on
- * Multirange limit 2
0 d ... 9999999 d, crossover from mean to large stepwidth
*only if Multirange mode = on
- Check
Counters for Calibration and Configuration (cannot be modified/reset)
- View
-
Same elements as
Param
(view only)
- Stepwidth
(view only)
- Fullscale
(view only)
- Deadload (mV/V)
(view only)
- Resolution (cd/c)
(view only)
- Calibrated at
(view only)
- Span (mV/V)
(view only)
- Sensitivity (µV/d)
(view only)