PR 5610 Installation Manual Commissioning with Terminal
Sartorius 113
The parameters for the serial ports are saved in EAROM.
♦ List of parameters and possible entries (selection):
• Selecting the serial interface: ( Serial IO ): 1, 2, … 7
• Selecting the previous/next serial interface:
Previous / Next serial I/O
• Protocol
: None XON/XOFF, DUST3964R slave, DUST3964R master, DUST3964R S5 slv, DUST3964R S5
mas, EW-COM Slave V1, EW-COM Slave V2, EW-COM Slave V3, RTS/CTS, JBUS/MOD-Bus, Remote
Display, W
&
M Printer, Modem, xBPI
• Number of data bits (
Bits
):
7, 8
• Parity check (
Parity): None, Odd, Even
• Number of stop bits (
Stopbits): 1, 2
• Transfer rate (
Baudrate
):
300, 600, 1200, 2400, 4800, 9600, 19k2, 38k4, 115k2.
• PR 5610 slave address
(Slave-adr): <space>, A to Z
• Terminal type (Devicetyp): Raw, Ansi, VT52
• Echo
(Echo): Disabled, Enabled
If elements appear in brackets on the screen, selection is not possible.
⇒ A PR 5610 can be equipped with max. 7 serial interfaces. For a survey, see chapter 5.2.3.1.
⇒ Dependent of selected protocol (and interface), defined default settings are made by the system.
5.2.3.1 Survey: physical interfaces
The instrument can besides the built-in interface be equipped with max. 6 additional serial interfaces, provided
that the 3 options sockets are fitted with 1713/04. The interface no. is allocated from TTY1 to TTY6 via software.
All interfaces can use each protocol, but EWCOM, Remotedisplay, xBPI and W&M printer only once per
instrument.
The following settings are useful dependent of protocol. These settings can also be changed subsequently.
Protocol Default settings Example Rem.
None
XON/XOFF 8, none, 1, 9600, -, ANSI, enabled System console 1
EWCOM slave V1 7, even, 1, 9600, A EW protocol V1 5, 6
EWCOM slave V2 8, even, 1, 19.2, A EW protocol V2, PR 1740 6
EWCOM slave V3 8, even, 1, 19.2, A EW protocol V3, DDE-server 6, 7
RTS/CTS standard: as XON/XOFF Printer 3
W&M printer 7, even, 1, 9600, -, raw I/O, disabled W&M printer 2
Remotedisplay 7, even, 1, 9600, -, raw-I/O, disabled PR 16xx, remote display 4
xBPI 8, odd, 1, 9600, Sartorius scale
Remarks:
1. A terminal or a PC with a terminal emulation program is used as system console.
2. An FX880/PH02 printer permits W&M weight data print-out.
3. Via this protocol, print-out of all configuration data and weight print-out during weighing point calibration
(if necessary) are made.
4. Thee instrument sends the remote display telegram successively for all 2 weighing point modules. This is
followed by a pause of approx. 80 msec.
Data contents: see PR 1627 and PR 1628 documentation.
5. Protocol for communication with PR 16xx units
6. For operation with PR 1750.
7. For operation with PR 1791 (DDE server).