PR 5610 Installation Manual Commissioning with Terminal
Sartorius 129
The system offers the following replies:
Unload and hit 0
unload the scale and set the zero
S to skip
skip zero setting
Q to quit
leave the calibration without doing anything
E to end
calibrate and finish
X to set fixed span
set the fixed span
P to Print
print out the weight value
2. Set the zero
Unload the scale, press key [0].
The instrument evaluates the data. The command line of the weighing point calibration mask is:
Setting deadloads …
the system stores the data. This is followed by:
Setting deadloads … Status OK
message is displayed during approx. 2 seconds
The following weight value is displayed:
0000 kg
Repeat zero setting in case of error. See section 3 and 5
3. The following command line is displayed in the mask:
Unload/Load
Calibration by weight
Unload and hit O or S to skip,
E to end, X to set fixed span, P to print
The system offers the following replies:
Unload and hit 0
unload the scale and set the zero
Load and hit Return
load the scale and set the span
E to End
finish the calibration
X to set fixed span
set the fixed span
P to print
print out the calibration parameters
4. Calibrating the weighing point
1. Place the calibration weights on the scale (if possible = full scale)
2. Press enter key [
]
The following line is displayed in the weighing point calibration mask:
Enter calibration point [ 6000kg ], E to exit:
The full scale is proposed as calibration weight.
3. Enter the calibration weight
Weight value with comma (or point), as applicable
- no or one or several spaces
- dimension (g, kg, lb or t)
- Press the enter key to complete [
]
The following command line is displayed in the weighing point calibration mask:
Enter calibration point [ 6000kg ], E to exit: 4000 kg
If the values are within the permitted range, the following message is displayed after a few seconds:
Setting unit … Status OK
Setting weight … Status OK
Message is displayed during approx. 2 sec.
Possible error messages:
Setting weight … Bad Weight
e.g. dimension missing
Setting weight … Status Above Scale
specified weight to high
Setting weight … Status Above phymax
input voltage > 36 mV
5. The command line as shown in point 3. is displayed:
Unload/Load
Calibration by weight
Unload and hit O or S to skip,
E to end, X to set fixed span, P to print
Repeat the calibration step (as shown in section 4.).
Apply the calibration weight
Load and hit Return
Press the Enter key [] (see 4.)