PR 5610 Installation Manual Hardware Configuration
Sartorius 23
Connection of one load cell in 4-wire technique:
• Connect + Supply (3) to + Sense (4) and - Supply (6) to - Sense (5) at the terminal block.
terminal LC connection description
1
2
3
4
5
6
7
+ Meas.
- Meas.
+ Supply
+ Sense
- Sense
- Supply
GND
+ signal / LC out
- signal / LC out
+ supply / excitation
+ sense
- sense
- supply / excitation
ground / screen
Connection of PR 6221 load cells
Please refer to the instruction manual PR 6021/08, -/68.
3.3.1.1 External supply for load cells