123
Continued on next page
Commands relating to calibration
Command Function
ECAL Starts external calibration
ECAL.W = XXX.XXX
Sets the reference weight value (W ref) for
calibration
XXX.XXX: Setting value
ETEST Starts an external calibration check
ICAL Executes calibration with the internal weight
ITEST Executes a calibration check with the internal weight
Commands relating to zero / taring
Command Function
ZRNG = X.XXX Sets the zero range X.XXX: Setting value
Commands relating to unit registration
Command Function
g Sets gram units
mg Sets milligram units (only accepted by models capable of displaying 0.001 g)
kg Sets kilogram units
ct Sets carat units
mom Sets momme units
CUo Sets / cancels user-specied units
o 0: Cancel, 1: Set
UCOFF = X.XXXX
Sets the conversion factor for user-
specied units
X.XXXX: Setting value
UDIG = X.XXX
Sets the minimum displayed value for
user-specied unit conversion
X.XXX: Minimum displayed
value
Other companies' commands
Command Function
TI Immediate taring (Mettler)
S Single output at stability (Mettler)
SI Immediate single output (Mettler)
SIR Continuous output (Mettler)
SR Continuous output at stability (Mettler)
(ESC) P Immediate single output (Sartorius) ESC = &H1B
(ESC) T Immediate taring (Sartorius) ESC = &H1B
Others
Command Function
"_" (space) Buffer clear command
# = XXXXXXX Enters and displays a numerical value XXXXXXX : Numerical value
{oo ... Echo-back mode
oo ...: Character string

Table of Contents