CALIBRATION
In the TW series, the weight for calibration is built in. The internal weight itself is calibrated on shipment
from the factory, but it is possible to recalibrate it using external weights. This is called
For the range of values for the external weights that can be used, refer to "Specications" (^ page 145).
1
Press for about 3 seconds.
This opens the calibration menu.
2
Select calibration of the internal weight.
[CAL.EXE] [P.CAL]
3
Enter the administrator’s password.
^
"Entering Numerical Values", page 49
^
"Changing the Password", page 141
4
Press
The password is acknowledged and the reference
weight value for calibration ashes.
*
If "WAIT" is displayed…
The calibration record is being output.
When output has nished, span
calibration will start automatically.
If the password is wrong…
The error message shown to the right will be
displayed and the display will return to the
indication in step 1.
!P!-!0!0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
!P!-!0!0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
!2!0!0.0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
!2!0!0.0!0!0
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
/
! !E!R!R! !N ! !E!R!R! !N
Calibration of the
Internal Weight
(TW Only)