Installation
1
BEFORE WEIGHING
TW/TX Series (small pan model) TW**3L, TX**3L
TW/TX Series (carat model) TWC**3L, TXC**3L
1
Fit the underplate.
Open the glass door and insert the underplate
slowly while tilting it.
Be careful not to knock against the surroundings.
2
Fit the pan ring.
(TWC**3L, TXC**3L only)
3
Place the pan supporter.
4
Place the pan on the pan supporters.
Align the two pan notches with the left and right
on the balance main body.
TXB Series (common to large pan models and small pan models)
1
Fit the pan ring.
Align the two pan ring notches with the
left and right on the balance main body,
and engage the four projections on the
pan ring in notches in the balance main
body.
Turn the pan ring counterclockwise until
it clicks into place.
Caution
Instructions
Turn the pan ring until it
clicks into place.
If the pan ring is not turned
sufciently, it will come into
contact with the pan, and the
display of the balance will become
unstable.
1
2
UnderplateUnderplate
Pan
Pan supporter
Pan ring
Pan
Pan supporter
Pan ring
Notches in the
pan ring
Notches in the
pan ring