EasyManuals Logo

SICK CLV 45 Series User Manual

SICK CLV 45 Series
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #90 background imageLoading...
Page #90 background image
Chapter 8 Operating Instructions
CLV 45x Bar Code Scanner
8-4
©
SICK AG · Division Auto Ident · Germany · All rights reserved 8 009 139/K949/06-06-2002
Troubleshooting
"
No valid code found
"
The Auto Setup was not suc-
cessful
No bar code was presented to
the CLV or the reading quality did
not reach 75 %
Repeat the Auto Setup.
Ensure that the bar code lies
within the reading range of the
CLV.
"
Oscillating mirror faulty
The CLV cannot operate the
oscillating mirror or cannot
initialize the mirror
Defective device Contact the SICK Service depart-
ment
"
Slave string faulty
" Data transfer error between a
slave and the master
Data connecting cables interrup-
ted
If the error occurs repeatedly,
check the data connecting cables
between all CLVs.
"
Wrong number of slaves
" During the master timeout,
the master receives more or
fewer reading results from the
slaves than expected from
the number of slaves entered.
Parameterized number of sla-
ves does not agree with the
number of connected slaves
Master timeout too short
Data connection between the
slaves interrupted
Check the entered number of
slaves. To do so, select upload
from the CLV by using the CLV-
Setup user interface, D
EVICE
CONFIGURATION
tab card and
check the number of slaves in
the CLV A
RRANGEMENT
section.
Download any changes to the
CLV!
As above, but check the time-
out defined for the master
Check the connecting cables
"Wrong parameters - master/
slave"
After selection of the master/
slave function, individual
parameter values, which are
not compatible with the auto-
matic settings defined for this
function by the CLV, were
changed subsequently.
Deviations from the following
values:
Test string: no (slaves)
Auxiliary input: no (slaves)
Pulse: Serial interface
or switching input
RK 512: without reaction
message
Use the CLV- Setup user interface
to check that the settings on the
D
ATA
STRING
,
T
ERMINAL
INTERFACE
,
D
EVICE
CONFIGURATION
,
and H
OST
INTERFACE
tab cards are
correct.
Message Meaning Possible cause Remedy
Table 8-1: Error messages output via the terminal interface

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK CLV 45 Series and is the answer not in the manual?

SICK CLV 45 Series Specifications

General IconGeneral
BrandSICK
ModelCLV 45 Series
CategoryBarcode Reader
LanguageEnglish

Related product manuals