5.4.1.1 Firmware
The FLOWSIC600 firmware is stored on a non-volatile memory (FLASH PROM). The pro-
gram code for the signal processor and system micro-controller are identified by a version
number (Reg. #5002 "FirmwareVersion") and a check sum (Reg. #5005 "ProgramCRC")
and can be verified as mentioned above.
NOTICE: Type approval
If the FLOWSIC600 is used for fiscal metering, the approved firmware versions
and the associated check sums are documented in the national pattern
approval certificates.