2.8 Data handling in the FLOWSIC600
2.8.1 Integrated volume counters
The FLOWSIC600 is equipped with integrated volume counters which can be displayed
both on the LCD display and in MEPAFLOW600 CBM.
Integrated volume counters
Last hour/day registers
Additional counters in meters with integrated Electronic Volume Corrector (EVC)
Mass counters
Volume counter Abbreviation
Volume at flowing conditions (forward) + Vf
Volume at flowing conditions (reverse) - Vf
Error volume at flowing conditions (forward)
1
+ Ef
Error volume at flowing conditions (reverse)
1
- Ef
Total volume at flowing conditions (forward) + Vo
Total volume at flowing conditions (reverse) - Vo
Total volume at flowing conditions (all) Vo
Volume counter Abbreviation
Forward volume of last hour Last hour forw.
Reverse volume of last hour Last hour rev.
Forward volume of last day Last day forw.
Reverse volume of last day Last day rev.
Volume counter Abbreviation
Volume at base conditions (forward) + Vb
Volume at base conditions (reverse) - Vb
Error volume at base conditions (forward)
1
+ Eb
Error volume at base conditions (reverse)
1
- Eb
Mass counter Abbreviation
Mass counter (forward) + M
Mass counter (reverse) - M
Mass total (forward) M+
Mass total (reverse) M-
Error Mass (forward)
1
1
All error volume counters can be reset (User Access Level "Authorized operator",
ď‚®
pg. 233,
9.3.5). The value of the error volume counter at the time of reset is stored in a logbook
entry.
Me+
Error mass (reverse)
1
Me-