44 FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Product Description
Subject to change without notice
Table 10 Warning preconditions
Configurable precondition Default value Notes
Warning Duration and
averaging for warnings
30s
The User Warning status becomes active only if a User Warning has been
continuously exceeded for the time specified in this parameter. All
measurements monitored for the User Warnings are averaged over the time
specified in this parameter.
Min. VOG for warnings
1 m/s
(3.3 ft/s)
The monitoring of measurements for the User Warnings only takes effect, if
the Avg. VOG is above the values specified in this parameter.
Caution: This parameter also defines the lower limit of the gas velocity range
classes for the Diagnostics Comparison Log . Changes to this parameter will
clear all data from the Diagnostics Comparison Log. This parameter is the
only User Warning parameter that can only be changed if the meter is in
Configuration Mode and in the User Access Level "Service".
SOS profile rising 0%/s
The SOS deviation warning only takes effect if the the SOS gradient between
all paths is below the limit specified in the parameter "SOS profilerising".