Commissioning
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG 155
Subject to change without notice
5.7.2 Configuration of DataLogs
5.7.2.1 Using the DataLogs
Starting with firmware version 3.4.00, the FLOWSIC600 provides two DataLogs (Hourly Log
and Daily Log). They save averaged measured values and are stored in the SPU‘s non-
volatile memory (FRAM). All data can be downloaded and exported to Excel files with
MEPAFLOW600 CBM (
pg. 53, 2.8.4.).
Valid flow for DataLog logging
Gas flow values are averaged for the DataLog entry only if VOG is above the
"Min. VOG for warnings" (
pg. 149, 5.7.1) and the flow is in the direction for
which the DataLog was configured. The Flow Time saved for every entry
indicates about how long the gas flow was valid for DataLog Logging during the
storage cycle. For example:
If the flow was above "Min. VOG for warnings" and in the forward direction
(for a forward configured Hourly Log) for half an hour, the corresponding
Hourly Log data will show a Flow Time of 50%.
If the flow was above "Min. VOG for warnings" but in the reverse direction
(for a forward configured Hourly Log) for an hour, the corresponding Hourly
Log data will show a Flow Time of 0%.
If the Flow Time is 0%, all diagnosis related values are shown as 0
(Temperature is 0 Kelvin). Meter Status and Volume Counter values are
shown as usual.

Table of Contents

Other manuals for SICK FLOWSIC600

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the SICK FLOWSIC600 and is the answer not in the manual?

SICK FLOWSIC600 Specifications

General IconGeneral
BrandSICK
ModelFLOWSIC600
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals