Maintenance
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG 181
Subject to change without notice
6.4.3 Diagnostics Comparison Report Check
To compare the current diagnostic values (current fingerprint) of the meter with the
reference diagnostic values (reference fingerprint, e.g. stored at commissioning), the
Diagnostics Comparison Report provides graphs and color coded information on
diagnostics.
Figure 86 Diagnostics Comparison Log table
To print a Diagnostics Comparison Report, complete the following steps:
Download the Diagnostics Comparison Log from your FLOWSIC600 according to
6.4.2.1.
Go to the Diagnostics Comparison data tab (
Figure 86).
Click the button "Preview/Print".
In the "Preview / Print options", enter your name and select the flow direction for which
you want to view the report.
Click "Preview / Print".
In the print preview, either print the report or export it to a PDF file.