EasyManua.ls Logo

SICK FLOWSIC600

SICK FLOWSIC600
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
78 FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Flow Calibration
Subject to change without notice
3.3.5 Parameter write lock
To calibrate the FLOWSIC600, the Parameter write lock has to be in the UNLOCKED
position.
Check the setting of the Parameter write lock:
Click the "System" button in the MEPAFLOW600 CBM main system bar (top right cor-
ner). The "Meter status" window is opened (
Figure 25).
Figure 25 "Meter status" window
If the parameter write lock status is LOCKED, it must be manually unlocked to enable
parameter changes (
Figure 26):
Disconnect the FLOWSIC600 from the power supply.
Take off the front cap of the signal processing unit.
Move the parameter write lock to the desired position.
Mount the front cap again.
Reconnect the power supply.
Figure 26 Parameter write lock
Parameter write lock status
(LOCKED or UNLOCKED)
Opens the "Meter
Status" window
Parameter write lock
LOCKED
UNLOCKED

Table of Contents

Other manuals for SICK FLOWSIC600

Related product manuals