Maintenance
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG 173
Subject to change without notice
6.2.4 Battery lifespan / capacity
The Real Time Clock (RTC) of the FLOWSIC600 is buffered by a battery. The manufacturer
states that the battery life span is at least ten years. The remaining battery capacity can be
viewed on the LCD in the first menu level (
pg. 222, 9.3.2).
Figure 82 Display of remaining battery capacity on the LCD display
Because the FLOWSIC600 has no regular maintenance cycle, the system alarm "Battery
lifetime" is generated when the remaining battery life is less than 15%. This alarm forces
the operator to change the battery (
pg. 189, 7.2.1). A logbook entry is also generated.
Optionally, the user can choose an additional user warning for the battery lifespan
(
pg. 150, Figure 75).
Battery
Charge 100%
NOTICE:
The battery may only be changed by trained staff. See
pg. 197, 7.2.4 for
troubleshooting.