52 FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG
Product Description
Subject to change without notice
Table 14 Dataset type "Mass Flow Counters"
Unit
Description Abbreviation Metric Imperial
Date/Time when data was saved Date / Time Sec Sec
Total volume at flowing conditions
(forward or reverse)
Vo forward or
Vo reverse
m³ acf
Error volume at flowing conditions
(forward or reverse)
Ef forward or
Ef reverse
m³ acf
Total Volume at base conditions
(forward or reverse)
VB Total forward or
VB Total reverse
Nm³ scf
Error volume at base conditions
(forward or reverse)
Eb forward or
Eb reverse
Nm³ scf
Total Mass (forward or reverse)
Mass Total forward or
Mass Total reverse
t lbs
Error Mass (forward or reverse)
Me forward or
Me reverse
t lbs
Temperature Temperature °C °F
Pressure Pressure bar(a) psi(a)
Molecular weight Molar mass g/mol lb/mol
Density Density kg/m³ lb/ft³
Last Volume during storage cycle
(at flowing conditions)
Vo log cycle m³ acf
Last Volume during storage cycle
(at base conditions)
VB log cycle Nm³ scf
Last mass during storage cycle M Total log cycle t lbs
Meter Status (
pg. 179) Status - -
● The dataset type "Mass Flow Counters" does not contain values for "Pres-
sure", "Temperature", "Molecular weight" or "Density" for gas velocities
below the value for Vmin ("LowFlowCutOff", Reg. #7036).
● The "Flow time" value shows, the percentage of storage cycle time the flow
was above Vmin and in the flow direction specified for the DataLog.
● The dataset type Mass Flow Counter cannot be configured for bidirectional
use. If this dataset is necessary for a bidirectional application, it is recom
-
mended that it be used for the Hourly log and Daily Log, and configured
accordingly.