Product Description
FLOWSIC600 · Technical Information · 8010125 V 4.0 · © SICK AG 59
Subject to change without notice
Session file
When a session is exported, a session file is generated which contains all data pertaining
to the meter with which it was created and all event entries in relation to the session. The
exported session file can be saved to the hard disk, transferred (e.g. via e-mail) or
imported to another meter database.
Event entry
An event entry is an entry that documents an important event that occurred during a ses-
sion. It is saved within this session and can be viewed and accessed via the "Meter
explorer" page. A session entry can contain the following event entries:
Event entries contain the event type, the name of the event (relevant for records, reports
and parameter changes), a time stamp and a description of the event.
Event entry Description
Time stamp (begin / end) Time stamps are written at the beginning and end of each session.
Last parameter set
The parameter set documents the parameter settings at the end of
the session. This entry is written when the session is opened, is kept
up to date during the session and is closed when the session is
ended.
Settings
Every change of any parameter is documented with the name of the
parameter, its old value and its new value.
Operating mode change
Whenever the meter is switched from Configuration Mode to
measurement mode or vice versa, the change is documented.
Unit switch Imperial Units /
Metric Units
Every change made to the unit system settings is documented.
Diagnosis session
Every diagnosis session (
7.3, p.198) created by the user is saved in
a separate session file and can be accessed (played, exported or
deleted) from the Meter explorer if the software is in an online or
offline connection with the particular meter.
Measurement record
Every measurement record (
2.9.4, p.61) created by the user is
saved and can be accessed (played, exported or deleted) from the
Meter explorer if the software is in an online or offline connection
with the particular meter.
Maintenance report
Every maintenance report created by the user is saved and can be
accessed (printed, exported or deleted) from the Meter explorer.
Error volume counter reset Every reset of the error volume counters is documented.
Logbook reset The clearing of data from a Logbook is documented
DataLog formatted
Changes to the DataLog format (
pg. 48, 2.8.3), which result in the
clearing of all data from the DataLog are documented.
DataLog cleared The clearing of data from a DataLog (
pg. 55, 2.9) is documented.
Firmware update Firmware updates (see Service Manual) are documented.