EasyManua.ls Logo

SICK FLOWSIC600 - Page 70

SICK FLOWSIC600
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
70 FLOWSIC600 ¡ Technical Information ¡ 8010125 V 4.0 ¡ Š SICK AG
Flow Calibration
Subject to change without notice
Testing at several test pressures
The applicable pressure range of a meter can be extended by testing the meter at several
test pressures. In this case, the pressure parameters must be calculated as follows:
where
p
e, min
= 0.5 * p
e, test, min
p
e, max
= 2 * p
e, test, max
Then the pressure parameters in the meter must be configured as follows (in the exact
order given):

Connect to the FLOWSIC600 with MEPAFLOW600 CBM (

pg. 77, 3.3.4)

Open the "Parameters" page (select Meter / Parameters)

Switch the meter to Configuration Mode (select File / Configuration Mode)

Select the parameter Reg. #7041 ''Pressure(fixed)"

Enter the value calculated for pfix

Click "Write to meter".

Select the parameter Reg. #7700 ''pe.min"

Enter the value calculated for pe,min

Click "Write to meter".

Select the parameter Reg. #7701 ''pe.max"

Enter the value calculated for pe,max

Click "Write to meter".

Switch the meter back to Operation Mode (select Meter / Operation Mode)
NOTICE:
The calibration curve correction for the extended applicable pressure range
must be applied by the flow computer. The current operating pressure must be
applied by the flow computer.
p
fix
= p
e, test, min *
p
e, test, max

Table of Contents

Other manuals for SICK FLOWSIC600

Related product manuals