1-164 280C-009-002 REV B www.stryker.com
Return To Table of Contents
English
The GoBed
®
II MedSurg Bed, Model FL28C is suited for use in the electromagnetic environment specified below.
The customer or the user of the GoBed
®
II MedSurg Bed, Model FL28C should assure that it is used in such an
environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment - Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the GoBed
®
II MedSurg Bed, Model FL28C,
including cables, than the recommended
separation distance calculated from the
equation appropriate for the frequency of the
transmitter.
Recommended Separation Distance
d=1.2
d=1.2
d=2.3
800 MHz to 2.5 GHz
where P is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
recommended separation distance in meters
(m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity of
equipment marked with the following symbol:
Note 1
At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the GoBed
®
II MedSurg Bed, Model FL28C is used
exceeds the applicable RF compliance level above, the GoBed
®
II MedSurg Bed, Model FL28C should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
reorienting or relocating the GoBed
®
II MedSurg Bed, Model FL28C.
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.
GOBED
®
II MEDSURG BED, MODEL FL28C (CONTINUED)
EMC Information

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker GoBed II and is the answer not in the manual?

Stryker GoBed II Specifications

General IconGeneral
BrandStryker
ModelGoBed II
CategoryMedical Equipment
LanguageEnglish

Related product manuals