www.stryker.com 280C-009-002 REV B 1-19
Return To Table of Contents
English
Problem / Failure Recommended Action
Improper operation of the CPR positioning:
Gatch does not lower and/or the Fowler actuator
does not reset.
• Check the two CPR limit switches and replace if needed.
The weight shown on the scale display (optional)
is incorrect.
• Zero the bed without the patient on it (see the “Scale System
Usage” section in the Operations Manual).
• Calibrate the scale.
Scale display (optional) reads OVERLOAD or
shows incorrect weight.
• The weight on the bed is greater than the scale capacity.
• If the patient weight is greater than 500 lb (227 kg), do not use
the scale. If that is not the case, zero the scale without the
patient on the bed.
• Ensure equipment added to the bed was done using the “Add/
Remove Equipment Procedure” (see the Operations Manual).
• One or more load cell cables are not properly connected or
completely disconnected from the scale control board (see
Figure 13, page 1-51) for the connecting positions of the load
cell cable).
Angle shown on the scale display (optional) is
incorrect.
• Calibrate the scale.
Scale display (optional) does not turn on when
the scale is activated.
• Verify the bed has power.
• Press ENTER key twice.
• Unhook the foot board from the bed by lifting it off the connec-
tors. Ensure the connectors match correctly and press the foot
board down firmly to reconnect it to the bed.
• Check the Scale user interface cable connection to the scale
control board (J6 connector, see drawing OL280005L, page
1-76).
Scale (optional) displays “No scale found”. • Press the ENTER key. If message persists, contact our Technical
Service. If message disappears, data displayed is correct and
current procedure can be resumed.
All lockout LEDs (padlock icons) flash or
“ERROR” appears at the bottom of the LCD
display.
• The scale is unable to measure the weight due to an electrical
problem. Please contact our Technical Service.
Bed Exit system (optional) does not operate
properly.
• Verify the load cell cable connections to the scale control board
(see Figure 13, page 1-51) for the load cell cable connection
positions).
Bed Exit system (optional) does not work at all
or alarm goes off when Bed Exit is activated.
• Check the load cell cable connections to the scale control board
(see Figure 13, page 1-51) for the load cell cable connection
positions).
Nurse call (optional) or Bed Exit (optional) signal
does not reach the nurse desk.
• Verify the bed has power.
• Check the connecting cable.
• Check the wall outlet.
• Contact the Technical Service.
Troubleshooting Guide

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker GoBed II and is the answer not in the manual?

Stryker GoBed II Specifications

General IconGeneral
BrandStryker
ModelGoBed II
CategoryMedical Equipment
LanguageEnglish

Related product manuals