FOOT END LIFT MOTOR REPLACEMENT (CONTINUED)
14. Reverse the above steps to install the new motor.
15. Plug the bed into the wall receptacle.
16. Lower the bed, and hold the bed down button until the lift motor reaches its lowest position.
CAUTION
The limits of the new motor must be adjusted prior to reconnecting it to the lift lever. An improper adjustment can damage
the lift mechanism.
17. To adjust the limits of the new motor, proceed as follows:
A. Ensure the motor cable is connected to the control board and plug the power cord into the wall receptacle.
B. Press the bed down control until the motor stops. This will be the lower limit of the motor.
C. Attach the molded nut support on the motor to the lift lever arms.
D. With the alignment fixtures still in place, raise and lower the bed completely to ensure the lower limit is
properly adjusted.
18. Test the Bed Up and Bed Down functions for proper operation before returning the bed to service.