Delay between pressing the <Start> key and the start of
measurement integration.
Synchronization of the measurement/integration start to
the nearest full minute or hour of the instrument real-time
clock. It helps to measure in full cycles.
Infinitive averaging of Summary Results that lasts from
start of the measurement till its cancellation with the
<Stop> key, remotely or if number of splits of the logger
file exceeds Split Cycles.
Maximum number of splits of the data registration in
separate files.
Time of averaging of Summary Results: from 1 second
to Infinitive. For example, with 8 hours integration period
the LEQ result will be averaged for 8 hours. In case of
Infinitive, the measurement will last until the user presses
the <Stop> key.
Number of measurement/integration repetitions after the
<Start> key pressure.
This enables to make a series of measurements without
pressing the <Start> key and save this series in the
results file.
Type of integration of RMS based results (RMS
detector): Linear or Exponential. The IEC 61672-1:2013
standard requires Linear integration, without time
weighting, however in some countries old regulation
refers to the Exponential RMS integration with standard
time weighting: Fast or Slow.
Linear type of integration of RMS based results (RMS
detector), without time weighting according to the
IEC 61672-1:2013 standard.
Exponential type of integration of RMS based results
(RMS detector), where averaging is a continuous
process that weighs current and past data differently.
The amount of weight given to past data as compared to
current data depends on the exponential time constant.
In the exponential averaging, the averaging process
continues indefinitely.