150
SV 977D User Manual
Statistics
Method for calculation of statistics for RMS results: with
linear detector (Lin) or exponential detector (Exp), e.g.
Impulse, Fast or Slow time constant.
Chapter
4.1
Day Time Limits
Definition of the day and night periods required by local
standards: 6–18h and 7–19h. These limits are used for
the calculation of the L(den) function.
Chapter
4.1
Rolling Time
Time frame for the "Rolling Leq" calculation. The Rolling
Leq is presented as LR+<time frame>. For example, if
the Rolling Time is equal to 30 minutes, the appropriate
result will be named as LR30 and calculated each
second as Leq integrated during last 30 minutes.
Chapter
4.1
Profiles
Virtual broadband level meters, which calculate the set
of results with own weighting filter (Filter) and exponential
detector time constant (Detector).
Chapter
4.3
Filter
Weighting filter applied in the profile in accordance with
most applicable world standards:
- for sound measurements: Z, A, C, B, U, AU, LF,
- for vibration measurements:
o acceleration: HP, HP1, HP3, HP10, Wh,
o velocity: Vel1, Vel3, Vel10 and VelMF,
o displacement: Dil1, Dil3 and Dil10.
Chapter
4.3
Appendix C
Appendix D
Detector
Exponential RMS detector time constant applied in the
profile:
- Impulse, Fast or Slow for sound results like Leq, Lmax,
Lmin, SEL, LEPd, Lden, Spl, Ltm3 and LTeq;
- from 100ms to 10s for vibration results like RMS, MAX.
Chapter
4.3
Appendix D
D.1.2
Compensation
Digital filter that compensates some effect: Microphone,
Free Field, Diffuse Field, Outdoor Environment,
Outdoor Airport and Windscreen.
Chapter
4.5
Microphone
Digital filter that compensates the microphone inner
noise. It is switched on by default, however it is possible
to switch it off for electrical measurements (e.g.
laboratory calibration measurements).
Chapter
4.5
Free Field
This option should be selected for sound measurements
in the free field conditions. The microphone supplied with
SV 977D is designed for sound measurements in the free
field conditions.
Chapter
4.5