Type of trigger that starts the wave recording for
Recording Time on condition: falling value of the
Leq/RMS result (Source) integrated during
Trigger Period passes below the threshold value (Level).
Type of trigger that starts the wave recording for
Recording Time on condition: value of the Leq/RMS
result (Source) integrated during Trigger Period is
greater than the threshold value (Level).
Type of trigger that starts the wave recording for
Recording Time on condition: value of the Leq/RMS
result (Source) integrated during Trigger Period is lower
than the threshold value (Level).
Type of trigger that starts the wave recording for
Recording Time when the triggering signal appears on
the I/O socket of the instrument.
Type of trigger, enabling the wave recording start for
Recording Time every time the measurement starts. If
Integration Period is shorter than Recording Time, the
wave recording will be continued for additional
Recording Time.
Type of trigger, enabling the wave recording start for
Integration Period every time the measurement starts.
Measured result that is compared with the threshold level
for triggering (Level) measured in the first profile: Leq(1)
in the sound mode and RMS(1) in the vibration mode.
Threshold level of the trigger condition.
Time interval of checking the triggering conditions. This
parameter can be set as: Log. Step (same as
Logger Step value), 0.5ms, 100.0ms and 1s.