188
SV 977D User Manual - Appendixes
71.29 first profile Leq result in dB
60.67 20.0Hz spectrum band result in dB,
63.57 25.0Hz spectrum band result in dB,
(...)
51.75 16.0kHz spectrum band result in dB,
56.18 20.0kHz spectrum band result in dB,.
Example 2
We want to get the following results every second:
- measurement time (with seconds resolution),
- overload information,
- first profile Leq results,
- second profile Leq results,
- third profile Leq results,
and then import them into a spreadsheet program (e.g. Excel) where the decimal separator is the comma.
To do this, we send command:
#R,1:3:9:44,G,V,R:1,R:2,R:3;
In response, instrument sends:
#R,1:3:9:44,G:0,V,R:1,R:2,R:3;
After measurement start, instrument will send below data every second:
#R<HT>11:22:56<HT>0<HT>94,64<HT>97,05<HT>98,92<HT>;<CR><LF>
#R<HT>11:22:57<HT>0<HT>95,09<HT>97,73<HT>98,90<HT>;<CR><LF>
#R<HT>11:22:58<HT>0<HT>94,25<HT>97,34<HT>99,08<HT>;<CR><LF>
#R<HT>11:22:59<HT>0<HT>96,49<HT>98,27<HT>99,12<HT>;<CR><LF>
#R<HT>11:23:00<HT>0<HT>95,74<HT>98,01<HT>99,13<HT>;<CR><LF>
#R<HT>11:23:01<HT>0<HT>95,52<HT>98,08<HT>99,10<HT>;<CR><LF>
#R<HT>11:23:02<HT>0<HT>97,83<HT>98,77<HT>99,10<HT>;<CR><LF>
#R<HT>11:23:03<HT>0<HT>95,06<HT>98,95<HT>99,11<HT>;<CR><LF>
where:
<HT> horizontal tab sign (HT, ASCII value 9),
<CR> carriage return sign (CR, ASCII value 13),
<LF> line feed sign (LF, ASCII value 10).
After importing the data into the spreadsheet program, we get results grouped into separate columns: