SV 977D User Manual - Appendixes
Group name
#1
code
#7
code
#S
code
Code description
type (read only)
1 - FAT16
2 - FAT32
3 - FAT12
SD-card number of
sectors (read only)
NS
#7,NS; This function returns the number of
sectors (sector = 512 bytes) on the SD-card in the
format #7,NS,x; where x number of sectors
(in case of x = -1 SD-card is not available).
SD-card number of free
sectors (read only)
NF
#7,NF; This function returns the number of free
sectors (sector = 512 bytes) on the SD-card in the
format #7,NF,x; where x number of free
sectors (in case of x = -1 SD-card is not available).
PIC version (read only)
PI
x.xx - internal microcontroller firmware version
number
Hardboot version (read only)
VH
x.xx - hardboot program version number
Bootstrap version (read only)
VB
x.xx - bootstrap program version number
File system version (read
only)
FS
x.xx file system version number
Bluetooth internal software
version (read only)
BR
x.xx bluetooth version number
Real time clock settings
RT
Current real time clock settings in the format:
#7,RT,hh,mm,ss,DD,MM,YYYY; where
hh:mm:ss denote the time and DD/MM/YYYY
gives the date.
Temperature (read only)
TP
x - temperature*10 °C
Factory settings (write only)
CS
#7,CS; This function clears current setup (without
RS232 and Bluetooth).
Marker definition
CM
#7,CM,<no>,"<name>",<event>,<point>;
where:
<no> marker number ,
“<name>” marker name,
<event> 0 (Off) or 1 (On),
<point> 0 (Off) or 1 (On),
Marker status
MR
#7,MR; This function returns the status of all
logger markers in the format:
#7,MR,<state1>,<state2>,<state3>,<state4>;
where:
<state1> status of the 1
st
marker (0 or 1),
<state2> status of the 2
nd
marker (0 or 1),
<state3> status of the 3
rd
marker (0 or 1),
<state4> status of the 4
th
marker (0 or 1).
#7,MR,<no>; This function returns the status of
the selected logger marker in the format:
#7,MR,<no>,<state>;
where:
<no> marker number ,
<state> marker status,
External Power
EP
#7,EP; This function returns the parameters of