SV 977D User Manual - Appendixes
Table C.3.5. Linear operating ranges with the Wh filter
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 12.5 Hz sine
signal
Table C.3.6. Linear operating ranges with the HP, HP1, HP3 and HP10 filters
Linear operating ranges (RMS)
With the SV 80 (10 mV/ms
-2
) accelerometer sensitivity @ 79.58 Hz
Measured at the ref. environmental conditions, at 80 Hz sine signal
Table C.3.7. Peak for the sinusoidal signal 1 kHz, at reference conditions
Peak for the sinusoidal signal 1 kHz, at reference
conditions @ 1 kHz (0.0 dB calibration factor)
1.41 V “HP10” – weighting
4.99 V “HP10” – weighting
Note: For the signals with the crest factor n >1.41 upper measuring range of the RMS is reduced.
The valid upper limit can be calculated according to the below given formula:
An dB(re1µV) = A dB(re1µV) – 20log(n/√2),
where A is the upper limit for the sinusoidal signal in dB(re1µV)
Example: For the crest factor n = 10 the upper limit is A
10
= 120 dB(re1µV)
Measuring frequency range 0.35 Hz ÷ 22.4 kHz with the HP filter (-3 dB)
Measuring frequency range 0.35 Hz ÷ 44.6 kHz with the HPE filter (-3 dB)