SV 977D User Manual - Appendixes
Note: For the conformance acoustical tests with SA 277D, the Microphone compensation and
the Outdoor Environment or Outdoor Airport compensation must be switched on (path:
<Menu> / Measurement / Compensation).
Statement of performance
SV 977D working as the SLM with SA 277D meets requirements of IEC 61672:2013 for the Class 1 Group X
instruments.
Linear operating ranges with the Outdoor Environment filter
The starting point at which tests of level linearity shall begin is 94.0 dB for the frequencies specifies below. For
the Low measurement range and A weighting linearity test at 31.5 Hz, the starting point is 74 dB.
Table C.7.1. Linear operating ranges for the Low measurement range and the Outdoor Environment filter
(for the sinusoidal signal and microphone sensitivity 50 mV/Pa)
Table C.7.2. Linear operating ranges for the High measurement range and the Outdoor Environment filter
(for the sinusoidal signal and microphone sensitivity 50 mV/Pa)
Table C.7.3. Self-generated noise for different weighting filters
*) measured with the ST 02 microphone equivalent impedance 18 pF +/-10%, averaging time – 30 s