TROUBLESHOOTING
INTRODUCTION
The instrument is switched ON but
will not start.
• Check if power cord is plugged in properly.
•Useanotherappliancetocheckiftheoutletislive.
•Checkfusesandcircuitbreakers,replaceifnecessary(see" Technical Maintenance"
page4-13).
The logo appears on the LCD
screen,
however the main screen does not
appear.
• Theprogramcardisnotinsertedproperly.TurnthemainpowerswitchO F F, and
checkwhethertheprogramcardissecurelyinsertedinthecardslotontheright
sideoftheinstrument.ThenswitchONagain.
After turning the main switch ON,
the LCD screen remains blank.
Possibilityofmemoryerror.TurnthemainswitchO F F, wait1to2minutes,thenturn
themainpowerswitchONagain.
Mechanical operation is heard, but
no display appears on LCD screen.
CheckwhethertheLCDscreencontrastissetcorrectly.
Fluid leaks from the instrument. TurnthemainpowerswitchOFFandwipeofftheleakeduid.
Danger, electric shock
Disconnect thepower cordbefore opening the
instrument.Otherwise thereisariskofinjuryby
electricalshockandpossibledamagetotheinstru-
ment.
Note
Iftheinstrumentshows amalfunction,checkthe
following table.Ifthecorresponding itemisnot
foundortheprocedurelisteddoesnoteliminate
theproblem,contactyourSysmexservicerepre-
sentative.
• Othererrorsareindicatedbyabeepandames-
sagedisplayedontheLCDscreen.
• Ifanerroraffectsonlyaspecicanalysisresult,
itwillbemarkedbyaag.
6.1. GENERAL ERRORS, INSTRUMENT FAILURE
Wearprotectivegarmentsandgloveswhenwork-
ing.Aftercompletionofwork,washhands with
disinfectant.
Biological risks