6-4
1234578 6
TROUBLESHOOTING
INTRODUCTION
pocH-100
i
Revised February 2011
6.3. ERROR MESSAGES, POSSIBLE CAUSES AND ACTIONS TO RESOLVE THE ERROR
Vacuum Error
Possible cause Defective vacuum pump or air leakage in vacuum line.
Action to resolve the error
Checkifanyuidwascollectedinthetrapchamber.
Press"Execute"tocleartheerrorandperformtherecovery
process. If the pressure returns to normal, Auto Rinse and
back ground check are performed in sequence, and the
screen returns to the Main screen.
The error message will be displayed again if the pressure
has not returned to normal.
Press"Cancel"toreturntothepreviousLCDscreen.
Special notes, if error re-
mains
Not ready until the error is resolved.
Piercer MC Error
PinchV1 MC Error
PinchV2 MC Error
Syringe MC Error
Possible cause Abnormality in the controller for each driving motor.
Action to resolve the error Turn the main switch OFF, then turn it ON again.
Special notes, if error re-
mains
Not ready until the error is resolved.
Replace pocH-pack 65
Possible cause LevelinpocH-pack65/pocH-pack65XLisinsufcient.
Action to resolve the error
ReplacethepocH-pack65/pocH-pack65XLwithnewone.
• Replacewithanewreagentandpress"OK".Theerroris
cleared and the reagent replacement sequence is executed.
See"Replacereagent"page4-12.
Anerrormessagewillbedisplayedagain,ifthereagentre-
placement sequence was not completed correctly.
Special notes, if error re-
mains
Not ready until the error is resolved.
Waste C. Error
Possible cause
Cloggingofwastechamberandtubing.
Wastetubingiskinked.
Pinchvalve2malfunctions.
Pinchvalve2tubingisblocked.
Defectivepressurepumporairleakinpressureline.
Floatswitchdoesnotworkproperly.
Action to resolve the error
Checkthecloggingofwastechamberandtubing.
Checkwastetubingforkinks.
• Afterchecking,press"Execute"tocleartheerror,andthe
waste chamber drain sequence is executed.
Special notes, if error re-
mains
Not ready until the error is resolved.
Piercer MT Error
Syringe MT Error
Possible cause Operation of each driving motor is abnormal.
Action to resolve the error Turn the main switch OFF, then turn it ON again.
Special notes, if error re-
mains
Not ready until the error is resolved.
Drain Error
Possible cause The waste bottle is full.
Action to resolve the error
See"Disposewasteuid"page4-8.
•A f t e r checking,press"Execute"tocleartheerror,andthe
recovery sequence is executed.
Special notes, if error re-
mains
Not ready until the error is resolved.