CHAPTER 6 Operation
Sysmex XE-5000 Instructions for Use 6-29
Revised April 2007 EU
1. Make sure that the instrument is in READY status.
The READY LED should be lit.
2. Press the MANUAL key on the Main Unit panel keypad.
The Manual Sample No. Setting screen will appear on the
LCD.
3. Using the numeric keys, input the sample ID number. Or,
read the barcode using a handheld barcode scanner
(option).
4. When analysis mode is not set in manual mode, press the
key and change the set item to “Mode.” Then using the
and keys, set the manual mode.
5. Use the key to change the setting parameter to
Discrete, then use the and keys to set to any of the
following.
4 CBC+DIFF+NRBC
5 CBC+DIFF+NRBC+RET
6. Use the key to change the setting parameter to
“Sample”, then use the and keys to set to HPC.
7. After all the settings are completed, press the ENTER key.
The analysis preparation sequence is performed.
Information
HPC analysis is only performed in manual mode. It
cannot be performed in sampler, capillary or manual
closed modes.
Information
The IMI detector is used for HPC analysis. HPC
analysis is not possible if the discrete setting does not
include DIFF analysis.
READY LED
Next No.123456789012346
123456789012345
1 2 3 4 5
CBC CBC CBC CBC CBC
DIFF DIFF
NRBC NRBC NRBC
RET RET
Sample 1:Normal 2:HPC 3:Body Fluid
DP No. 123456789012345
C D N R
Num
DP
Xm
Manual Capillary Closed
1 2 3
Manual
Not Ready
<Select Mode and No.>
Mode
Discrete
Sample No.