EasyManua.ls Logo

SYSMEX XE-5000 - Pneumatic Unit

SYSMEX XE-5000
262 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CHAPTER 3 Design and Function
Sysmex XE-5000 Instructions for Use 3-11
Revised April 2007
7 Displays the input status by numeric keys.
Alphabet and hyphen in addition to numerals can be used for sample ID No.
By pressing the NUM./ALPH. key on the panel keypad, you can change the input mode
in this order: “Numeric” “Uppercase Alphabet” Lowercase Alphabet”. The display
contents are as follows:
Num: Numeric mode
ALP: Uppercase Alphabet
alp: Lowercase Alphabet
8 Displays the status of the DP.
DP: DP is connected and no error has occurred.
No indication: DP is not connected.
DP (reversing display): DP is set to be connected, but an error has occurred.
9 Displays the implementation status of X
M QC.
Xm: X
M QC is ON.
No indication: X
M QC is OFF.
10 Displays the function menu.
The contents of the function menu differ with each other.
3.3 Pneumatic Unit
Front View
1 Pressure Indicator 1 (0.25 MPa)
Indicates pressure to be supplied to the Main Unit. The normal range is 0.25 ±
0.03 MPa. If pressure comes out of the normal range, the instrument may not operate
properly.
2 0.25 MPa regulator
Regulates the 0.25 MPa pressure to be supplied to the Main Unit.
3 Vacuum Indicator 2
Indicates vacuum to be supplied to the Main Unit. The normal range is -0.05 MPa or
over. If vacuum is less than -0.05 MPa, the instrument may not operate properly.
4 Pneumatic unit trap chamber
Prevents reagent, etc. from flowing into the compressor when an abnormality has
occurred in the instrument.
1
2
3
4
5

Table of Contents

Other manuals for SYSMEX XE-5000

Related product manuals