CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-7
Revised April 2007
12. Turn ON the power of the Main Unit. Background check
will start automatically. Make sure that all background
values are within acceptable.
13. Make sure by quality control that the instrument has no
performance problem.
14. Reset the SRV counter.
See “Chapter 10: 10.5 Status”.
9.4 Cleaning the manual rinse cup
If blood adheres to the manual rinse cup or any clogging is
found, clean the manual rinse cup by the following procedure.
Information
• Place the rotary valve with the notch facing upward
and the metal knob positioned between the stoppers.
• Confirm the metal knob is placed between the
stoppers.
If it is not, malfunction can occur.
• Make sure that the manual rinse cup is pushed all
the way up with the manual aspiration pipette
inserted. If the power is ON with the rinse cup
staying lowered, “Rinse Motor Error” will occur,
making it impossible to continue the operation.
Information
If you spill reagents on the coated surface of the
instrument, immediately wipe it off using wet cloth or
the like.
Groove
Stopper
(Viewed from the left side)
Metal knob
Guide pin for
rotary valve
Risk of infection
When cleaning the manual rinse cup, always wear
gloves. After completion of the operation, wash your
hands with disinfectant.
If your hands are contaminated by blood, you might be
infected by pathogen, etc.