CHAPTER 9 Cleaning and Maintenance
9-8 Sysmex XE-5000 Instructions for Use
Revised April 2007
1. Turn OFF the power of the Main Unit and Pneumatic Unit,
and wait several minutes until the pressure gauge points
at “0”.
2. Open the Main Unit front cover.
3. Gently push down the manual rinse cup with both hands.
Make sure that the manual rinse cup is completely
removed from the manual aspiration pipette.
4. Remove the manual rinse cup in the order of (1), (2) and
(3).
5. Wash the manual rinse cup with running water.
6. Make sure that no dirt remains in the manual rinse cup
and wipe off moisture.
7. Assemble the manual rinse cup in the reverse order of
disassembly. Install so that the small tubes pass behind
the manual rinse cup.
8. Gently push up the manual rinse cup all the way using
both hands.
Warning!
Before operating with the front cover open, be sure to
set the stop bar. Otherwise, the cover can drop down
and injure your head.
Manual aspiration pipette
Manual rinse cup
Manual rinse cup
(1)
(2)
(3)