CHAPTER 4 Reagents
4-4 Sysmex XE-5000 Instructions for Use
Revised April 2007 EU
4.8 STROMATOLYSER-NR(S)
Intended Use
Dye solution for the NRBC channel of hematology analyzers.
Storage and Shelf Life after first Opening
Store STROMATOLYSER-NR(S) at +2 to +35°C and out of
direct sunlight.
Do not use reagent that may have frozen.
If the container is unopened, it may be used up to the expiry
date stated on the reagent container.
Please refer to the product labeling (package insert or outer
package) for the open stability.
Usage
STROMATOLYSER-NR(S) is a reagent using in combination
with STROMATOLYSER-NR(L) to analyze erythroblast count
and ratio by the flow cytometry method, using a
semiconductor laser.
4.9 SULFOLYSER
Intended Use
A cyanide-free lyse reagent for the hemoglobin determination
in hematology analyzers.
Storage and Shelf Life after first Opening
Store SULFOLYSER at +1 to +30°C and out of direct sunlight.
If it has frozen, thaw it in warm water at 30°C or below, and
mix it thoroughly before use.
If the container is unopened, it may be used up to the expiry
date stated on the reagent container.
Please refer to the product labeling (package insert or outer
package) for the open stability.
Usage
SULFOLYSER is a reagent used in combination with the
specified diluent to analyze the hemoglobin concentration
dissolved out of RBCs by the SLS-hemoglobin method.