CHAPTER 2 Safety Information
2-2 Sysmex XE-5000 Instructions for Use
Revised April 2007
2.3 Installation
Warning!
• Never put the power plug in any socket other than the
specified voltage. When installing the instrument, be
sure to ground it. Otherwise, fire or electrical shock
may result.
• Take care not to damage the power cord, not to
place a heavy device on it and not to pull it forcibly.
Otherwise, the wire may break causing fire or
electrical shock.
• When connecting the instrument to a peripheral
(host computers and printer etc.), be sure to turn off
the power switch beforehand. Otherwise, electrical
shock or instrument failure may result.
Caution!
• Install the instrument in a place which is not subject
to water splash.
• Install the instrument in a place which is not subject
to adverse effects of high temperature, high humidity,
dust, direct sunlight, etc.
• Do not give the instrument a strong vibration or
impact.
• Install in a well-ventilated area.
• Do not install near devices that cause signal noise,
such as radios and centrifugal machines.
• Do not install the instrument near a chemical storage
or a place where a gas is generated.
• Do not use this instrument in any operating
environment which has electro-conductive or
flammable gases, including oxygen, hydrogen and
anesthetic.
• This instrument was designed for indoor use only.