CHAPTER 9 Cleaning and Maintenance
9-2 Sysmex XE-5000 Instructions for Use
Revised July 2007
9.1 Execution of shutdown
By executing shutdown, the detector chamber and dilution line
are cleared. Execute shutdown every 500 samples or at the
end of each day’s analysis, and also at least once every 24
hours when you use this instrument continuously.
1. Press the SHUTDOWN key on the Main Unit panel
keypad.
2. The shutdown screen appears on the Main Unit LCD
screen.
To cancel shutdown, press the RETURN key or select
“Cancel” on the function menu of the shutdown screen to
return to Ready status.
3. Set CELLCLEAN to the manual aspiration pipette as
shown in the figure; then press the START switch in that
status.
While the READY LED is blinking and the buzzer is
sounding, aspiration is in progress. Keep holding
CELLCLEAN in the current status.
4 After the READY LED turns off and buzzer stops, remove
the CELLCLEAN.
Note:
When the analyzed sample exceeds 500 samples in a
day, the message prompted to shutdown will appear.
Caution!
• Be sure to execute shutdown to clean the instrument.
Otherwise analysis may not be performed correctly.
• To execute shutdown, use CELLCLEAN only.
Warning!
CELLCLEAN is a strong alkaline detergent. Pay
careful attention not to let it adhere to your body or
clothes. If CELLCLEAN adheres to your body or
clothes, wash it away immediately with plenty of water.
Otherwise, it can damage your skin or clothes.
Note:
The manual aspiration pipette is automatically rinsed,
and does not need to be wiped manually.
Next No.1
DP No.
C D N R
Manual
Cancel
Not Ready
It will take approx. 15 minutes
Set CELLCLEAN to the pipette
and press Start Switch
CAUTION:Do not use detergents
other than CELLCLEAN.
Num
DP
Xm
READY LED
CELLCLEAN
START Switch