CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-41
Revised April 2007 EU
3. Remove three screws on the piercer fixing board and
remove the board.
4. Install the tubes and rubber joint which are connected to
the rinse cup and piercer.
5. Install the piercer cover and CP cover.
6. Close the front cover.
7. Reset the piercer cycle counter.
See “Chapter 10: 10.5 Status”.
9.17 Replacing the hand clipper
If the hand clipper is deformed and unable to hold the sample
tube, replace it in the following procedure:
Caution!
When installing the piercer cover and CP cover, take
care not to have the tube caught or bent.
Otherwise, analysis may not be performed correctly.
Piercer fixing board
Rubber joint
Rinse cup
Piercer
Tube
Risk of infection
When replacing the hand clipper, always wear gloves.
And after completion of the analysis, be sure to wash
your hands with disinfectant.
If your hands are contaminated by blood, you might be
infected by pathogen, etc.
Warning!
Before operating with the front cover open, be sure to
set the stop bar. Otherwise, the cover can drop down
and injure your head.