CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-43
Revised April 2007 EU
9.18 Replacing rubber plate No. 39
If the hand clipper becomes slippery, replace the anti-slip
rubber plate No. 39 in the following procedure:
1. Turn OFF the Main Unit power and open the front cover.
2. Loosen the fixing screws and remove the CP cover.
3. Remove the hand clipper fixing screws (2 places) and
remove the hand clipper.
4. Remove the rubber plate No. 39.
5. Install a new rubber plate in the reverse order.
6. Install the hand clipper to the original position.
7. Install the CP cover.
8. Close the front cover.
Risk of infection
When replacing the rubber plate No. 39, always wear
gloves. And after completion of the analysis, be sure
to wash your hands with disinfectant.
If your hands are contaminated by blood, you might be
infected by pathogen, etc.
Warning!
Before operating with the front cover open, be sure to
set the stop bar. Otherwise, the cover can drop down
and injure your head.
Caution!
When installing the CP cover, take care not to have
the tube caught or bent.
Otherwise, analysis may not be performed correctly.
Rubber plate No. 39
Hand clipper