CHAPTER 6 Operation
Sysmex XE-5000 Instructions for Use 6-17
Revised April 2007 EU
6.17 Sample analysis in manual mode
1. Make sure that the instrument is in READY status.
The READY LED should be lit.
2. Press the MANUAL key on the Main Unit panel keypad.
The Sample No. Setting screen will appear on the LCD.
3. Using the numeric keys, input the sample ID number. Or,
read the barcode using a handheld barcode scanner
(option).
4. When analysis mode is not set in manual mode, press the
key and change the set item to “Mode”. Then using the
and keys, set the manual mode.
5. Check the discrete setting, and if necessary, press key
to change the set item to “Discrete”. Then using the
and keys, set the discrete.
6. Use the key to change the setting parameter to
“Sample”, then use the and keys to set to Normal.
7. After all the settings are completed, press the ENTER key
to confirm.
Risk of infection
When analyzing samples, be sure to wear gloves.
After completion of the analysis, be sure to wash your
hands with disinfectant.
If your hands or body are contaminated by blood, etc.,
you might be infected by pathogen, etc.
Note:
Refer to “Chapter 6: 6.23 HPC analysis” when
performing HPC analysis, and to “Chapter 6: 6.24
Body fluid analysis” when performing body fluid
analysis.
READY LED
Next No.123456789012346
123456789012345
1 2 3 4 5
CBC CBC CBC CBC CBC
DIFF DIFF
NRBC NRBC NRBC
RET RET
Sample 1:Normal 2:HPC 3:Body Fluid
DP No. 123456789012345
C D N R
Num
DP
Xm
Manual
Not Ready
<Select Mode and No.>
Sample No.
Mode
Discrete
Manual Capillary Closed
1 2 3