CHAPTER 9 Cleaning and Maintenance
9-6 Sysmex XE-5000 Instructions for Use
Revised April 2007
8. Clean the surface of the fixed and rotary valves using
distilled water or 1:10 dilution of CELLCLEAN detergent.
Make sure to clean them with distilled water after cleaning
with CELLCLEAN.
9. Clean the removed tray in running water.
10. Make sure that the valve contact surfaces are thoroughly
free from dirt or dust.
11. Assemble the sample rotor valve in the reverse order of
disassembly. And gently push up the manual rinse cup
with both hands.
Note:
The valves are in close contact with each other. They
can be easily removed when you lightly pull, gently
twist, then slide them.
Caution!
Take care not to inflict flaws or scratches on the
sample rotor valve surfaces. Do not use each valve
with dust or dirt adhered on its surface.
These can cause blood leakage and incorrect analysis
results.
Information
Be sure to use CELLCLEAN only. Although the
sample rotor valve is corrosion-resistant against
CELLCLEAN, wipe it off completely to prevent
troubles to the unit or other components.