CHAPTER 9 Cleaning and Maintenance
Sysmex XE-5000 Instructions for Use 9-5
Revised April 2007
4. Gently push down the manual rinse cup with both hands.
Make sure that the manual rinse cup is completely
removed from the manual aspiration pipette.
5. Remove the constant-pressure screw.
6. Remove the sample rotor valve assembly.
7. Remove the rotary valve.
Information
If the manual rinse cup is not completely removed
from the manual aspiration pipette, there is a
possibility that the manual aspiration pipette may bend
when the sample rotor valve is removed.
Caution!
Do not pull out the valve excessively. This is to prevent
excess force from applying to the tube that is
connected to the rear fixed valve.
Information
• When removing the fixed valve, reagent can leak
from the tube. Have a paper towel ready to wipe
reagent.
• When pulling out the valve, take care not to bend the
manual aspiration pipette.
Manual aspiration pipette
Manual rinse cup
Constant-pressure
screw
Loosen
Sample rotor valve