Magnetic Balance
13.5.3 Data handling definitions
Storage:
Overwrite when a measurement is
repeated, results overwrite the actual
measured values for the tap
Append When a measurement is
repeated, results are added in the list but
previous results are also kept.
Recording:
Measurement can be done in auto stop mode
(value is automatically recorded when stable)
or in manual mode (unit stays measuring until
the operator click the stop button), selection
between both operations modes is done in
the setup screen, Data folder.
To record a value while measuring on this
mode, click the record button in the function
key bar.
13.5.4 Test Voltage Selection
Selection of the measuring voltage is done by
clicking on the test voltage field in the main
screen, any value between 0 and 67V can be
introduced, some values are directly
selectable.