EasyManua.ls Logo

Tettex 2293 - Exporting and Printing Results

Tettex 2293
226 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Short circuit impedance
151
Connection
Clamps used during the test
Short
Short circuited winding
Voltage
Voltage applied during the measurement
Current
Current applied during the measurement
Frequency
Frequency applied during the measurement
Impedance
Single phase impedance measured (Phase Voltage/ Phase Current)
Short circuit Imp ()
Single phase short circuit impedance calculation
After all phases are completed, then a calculation of the transformer overall values is done based
in the single phase measurements
Impedance
Calculated three phases Impedance
Short circuit Imp ()
Calculated three phases short circuit impedance
Voltage
Calculated three phases Voltage
14.4.10 Exporting and printing results
See the chapter “Data Handling” for information about exporting and printing results.

Table of Contents