Res:Set:Current LV,8.9
Res:Set:Current TV,13.2
Res:Set:Phase:Select HV,1,1,0
Res:Set:Phase:Select LV,1,0,1
Res:Set:Phase:Select TV,1,1,1
Res:Set:Winding:Select HV-LV
Res:Set:Data:Stability 0.0003
Res:Set:Data:Recording Auto
RemoteToLocal
After sending all the settings, a
Meas:Start
will start the resistance measuring sequence. After the measurement has started, the measuring
sequence and mode can be obtained using
Res:Meas:Sequence?
The result of this query helps to find out which winding and phase the device is currently
measuring. Further
Res:Meas:State
Shows which sequence step is currently performed and if a stable resistance reading is reached.
Meas:State
Has to be used to find out in which state within a sequence step the device currently is. For
resistance measurement the state flow will typically look like
Standby Initializing Charging Measuring Discharging
Reset
Standby
When “Data Recording” is set to auto, the device will automatically continue from Measuring to
Discharging. When “Data Recording” to manual, the device stays in the Measuring state until